Glucose readings from a sensor sampling in interstitial fluid aka Interstitial Glucose concentration [IGc] differ substantially from Blood Glucose Concentration [BGc] values measured at the same time.
Results show an average total lag of 17.2 ± 7.2 min for all subjects evaluated. The instrumental lag was 13.5 min, suggesting that physiological lag is approx 5 min. In addition, when glucose was increasing, the change in IGc was less than that in BGc, while when BGc was decreasing, the change in IGc was greater than that in BGc.
Differences reflect physiological variation in glucose uptake, utilization, and elimination in blood and interstitial space. This further evidence of the difference between IGc and BGc should be considered when interpreting glucose measurements.